Author: SIKAND Trimaan Singh

At LINNC Asia 2023, Trimaan Singh Sikand presents a unique case: a 63-year-old patient with a complex aneurysm in the internal carotid artery. During a flow diverter stent procedure, an unexpected device prolapse occurred. Swift rescue measures were taken, offering valuable lessons: choose wider and longer devices for better anchorage and consider coiling before full deployment for added support.